HOMETechnical SupportFlowmeter Calibration and Performance TestInfrastructure Facilities
Facilities
- Water meter performance test
- Calibration and Measurement Capability(CMC) : 0.08 ~ 0.15 %
- Scope of flow measurement : 0.01 ~ 40 ㎥/h
- Measurement method : Weight method, master meter method
- Testing diameter : 15 ~ 50 mm
Water meter performance test Table
Measurement method |
Device |
Scope of measurement |
CMC |
Gravimetric method |
Scale 600 kg 50 kg |
(0.01 ~ 40) ㎥/h |
0.08 % |
Master meter method |
Mass flow meter |
0.15 % |
- Calibration at Permanent Standards Laboratory (indoor)
- Total floor area: 2,600 ㎡
- Calibration and Measurement Capability(CMC) : 0.08 ~ 0.40 %
- Scope of flow measurement: 3 ~ 2,700 ㎥/h
- Measurement method: Weight method, master meter method
- Maximum diameter: 800 mm
- Collection tank: 1, 6, 10, 50 ton
Calibration at Permanent Standards Laboratory(Calibration method,Scope of measurement,CMC)Table
Calibration method |
Scope of measurement |
CMC |
Weight method |
(1 ton) 3~20 ㎥/h (6 ton) 20~200 ㎥/h |
0.20 % |
(10 ton) 200~500 ㎥/h (50 ton) 500~2,700 ㎥/h |
0.08 % |
Master meter method |
3 ~ 2,700 ㎥/h |
0.30 % |
open channel |
21~190 ㎥/h |
0.45 % |
- Constant pressure tank
- Height: 27.5 m
- Capacity: 68 m
- For preventing whirlpool and providing stable flow
- Mater meter
- High-precision reference flowmeter
- Measurement pipe : caliber
- between 50 and 800 mm
- Upper straight length 50D, lower straight length 10D (stabilizing liquid flow shape distorted by bend pipe and valve)
- Supply pump
- 4 150 kW units
- 1 75 kW unit
- Collection tank
- 1, 6, 10, 50ton Collecting fluid passed through hydrograph and measuring weight
- Underground collection tank
- 810 m (reservoir containing enough water in order not to make error by a change in water temperature during calibration)