물 자연 그리고 사람 - 물로 더 행복한 세상을 만들어가겠습니다.
HOME성과논문실적

논문실적

수질 분석 전처리 자동화 설비 효과성 평가(1) 게시글의 제목, 학술지명, 저자, 발행일, 작성내용을 보여줌
수질 분석 전처리 자동화 설비 효과성 평가(1)
학술지명 한국환경분석학회 저자 현승미,여상구,한승철,최재원,김윤석,이선홍,윤미애,이재희
발표일 2019-03-29

The automated measurement system for water quality analysis is customized. After the sample with the bar
code is introduced on the conveyor, it is divided into the aliquot and the online measurement section. The system
performs each for the sub-sampling of the original sample and the online measurement and data registration
process of pH, turbidity, electric conductivity (EC) and total dissolved solids(TDS). For sub-sampling,
a total of 17 group sample aliquots were carried out based on the criterion for drinking water quality test. In
terms of precision and accuracy, the results confirmed that the statistically equivalent data between automation
system and the human performances. The effectiveness of the automation equipment was evaluated by comparing
the unit time of the aliquot and online measurement. For drinking water quality testing, 0.32 hour unit
time were calculated by combining 0.17 unit hour per sub-sampling and 0.15 unit hour per online measurement.
The cost recovery period, which takes into consideration the facility price is expected to take 1.8 to
2.8 years. The automation equipment is expected to play a role as a cooperative robot to help lab worker,
because it will replace the sample management and the simple measurement if the installation cost is payable
according to the built and assemble in domestic and increased demand in the future.

목록