Development of real time monitoring of off-flavour compounds using multi functionalized auto sampler with SPME-GC-MS/MS 게시글의 제목, 학술지명, 저자, 발행일, 작성내용을 보여줌
Development of real time monitoring of off-flavour compounds using multi functionalized auto sampler with SPME-GC-MS/MS
학술지명 한국질량분석학회 저자
발표일 2015-08-20

Lately, needs of real time monitoring of off-flavours are increased due to the algal blooming in intake area for drinking water treatment plants(DWTP). Using real time data, rapid response to off-flavours by adsorption and oxidation treatments is important issue in DWTP. This study developed real time monitoring of off-flavours using multi functionalized auto sampler with SPME-GC/MS. 
Reference materials of geosmin, MIB and 1,2-dichlorobenzene D4 were purchased from Accustandard (USA). To real time based drinking water sample introduction to the GC/MS, PAL RTC auto sampler (CTC Analytics AG, Switzerland) was used. GC/MS by Bruker 300MS was combined to the real time sample introduction system. Due to the flexible function set of PAL RTC auto sampler, basic program for real time direct sample introduction and pre-treatment were simply set as following; 1) sampling and transport of 10mL drinking water from tap water pipe to the sample bottle with septa, 2) spiking internal standard by 10μL syringe, 3) mixing the sample, 4) heating, agitating and SPME adsorption, 5) desorption and GC/MS run. Syringes were sequentially changed by programmable adaptor system. Evaporative loss of target compounds during step 1) and step 3) was checked. Geosmin and MIB were detected within 18min with LOQ of 0.5 ng/L and 0.6 ng/L. Optimal run time of whole system was set at 2 hour. This means 12 data can be obtained in a day. 
Using this system, demonstrations have been operating from early June. Increased temporal trend was observed for geosmin during 2 months. This real time monitoring method provides 1) simple instrumentation, 2) harmonized with accredited method for drinking water in MOE including adding internal standard and salt

목록