Construction of K-water Model Turbine Test Facility
학술지명 AFORE2013 저자 김경성,탁세완,조용,오용환
발표일 2013-11-06

The reduced scale homologous hydraulic turbines are generally tested in order to confirm the performance of the prototype turbine. The model test facility should be able to reproduce the prototype operating conditions. The facility in K-water as shown in Fig. 1 is composed of a pump to make a pressure head, head and tailrace tanks to reproduce inlet and outlet of the turbine, a dynamometer to absorb power of the turbine, a heat exchanger to refrigerate water, a compressor and a vacuum pump to control pressure, many measuring equipment such as flow meter, pressure gage, and torque meter. Model test conditions for domestic hydraulic turbines have been investigated, and the facility has been designed to reproduce the test condition of head of 40 m and flow rate of 1.0 m3/s based on IEC 60193[1]. To stabilize the flow inside the pipeline, reducers, straighteners, long radius elbows, and partitions are installed in the right positions. Since the accuracy of the flow meter is closely related with the system accuracy, it is installed with enough distance for fully developed flow and flow stabilizing devices such as reducer and straightener. Also, the calibrated measuring equipment is installed at the right place and partial equipment is calibrated again on a site.

목록