습식다회선 초음파율야계의 특성평가 연구 |
---|
학술지명 한국유체기계학회
저자 박태진,고재명,이동근,조용,박종호
발표일 2013-08-01
|
An experimental investigation has been carried out in order to evaluate characteristcs of wetted-type multi-path ultrasonic flowmeters. The multi-path ultrasonic flowmeters were installed at various entrance and exit locations for several cases of pipe fitting(straighr, 90 double elbow) and valve(gate, butterfly). We measured the flow-rate at each location. The measurement data of test flowmeter were compared with the measured data of reference flowmeter. The uncertainties of reference flowmeter and test flowmeter are 0.3%, 0.4%, respectively. The results demonstrate the effects of flowmeter location as well as the measurement errors in flow rate. The distance between the flow disturbance factor and a flowmeter was an important element of the test. |