Nitrosamine Analysis by Semi-automated SPE and Fully Automated GC-MS/MS
학술지명 2013 APCE/KSFEA/APIA 저자 김윤덕,Huebsc,Hans-J,정중목,최재원
발표일 2013-11-05

Conventional nitrosamine analysis needs manual SPE sample preparation and GC-CI/MS/MS
method based on EPA method 521(2004). In this study, automated system for sample
preparation and GC-MS was optimized. GC-EI-MS/MS measurement was done by TSQ
8000 model with TriPlus auto sampler (Thermo Fisher Scientific). Calibration standards were
automatically prepared by using the advanced TriPlus CTC technique. For SRM (seltected
reaction monitoring) optimization, fully automated SRM was performed for 9 nitrosamines.
SRM optimization was done within a day by using automated SRM. In the normal condition,
initial instrumental detection limit (IDL) of NDMA was 3 pg. IDL of other nitrosamines were
2-3 pg ranges. This indicates the low ng/L range LOQs of treated water samples.
On the other hand, various brand carbon cartridges showed drastic recoveries. This was
compared using automated SPE system (GL Science). Actual operation of SPE system was
modified into three steps, 1) conditioning and loading 2) drying and elution, and 3) further
concentration. Spiking test using real treated water was performed by above system. In terms
of S/N ratio, method detection limit (MDL) of NDMA was 1.59 ng/L. MDLs of other
nitrosamines were between 0.08 and 0.61 ng/L, respectively. After minor modifications, the
method was used for nationwide nitrosamine monitoring.