Differences between Analytical Results for Perchlorate by Ion Chromatography-Conductivity detection and Ion Chromatography with Tandem Mass Spectrometry Considering EPA Method 314.1 게시글의 제목, 학술지명, 저자, 발행일, 작성내용을 보여줌
Differences between Analytical Results for Perchlorate by Ion Chromatography-Conductivity detection and Ion Chromatography with Tandem Mass Spectrometry Considering EPA Method 314.1
학술지명 Journal of Analytical Chemistry 저자 김학철
발표일 2011-11-30

In this study, an analytical method was developed for determining the perchlorate concentrations
in effluent samples obtained from two refineries. The primary and confirmatory methods used
were from EPA 314.1. The two methods found similar perchlorate concentrations for the Y
refinery sample; however, there was a significant difference in the concentration values found for
the K refinery sample. The difference was attributed to co-eluting ions with nearly the same
retention time as perchlorate. For more accurate and reliable analysis, the two samples were
analyzed by Ion Chromatography with Tandem Mass Spectrometry(IC-MS/MS) in Multiful
Reaction Monitoring(MRM) mode. The results from IC-MS/MS corresponded well to the
confirmatory method. In conclusion, these results show that the primary method of EPA 314.1
can return false data due to co-eluting ions with perchlorate.