물 자연 그리고 사람 - 물로 더 행복한 세상을 만들어가겠습니다.
HOME성과논문실적

논문실적

IC/Cond.와 IC/MS/MS를 이용한 수중의 Perchlorate 분석 게시글의 제목, 학술지명, 저자, 발행일, 작성내용을 보여줌
IC/Cond.와 IC/MS/MS를 이용한 수중의 Perchlorate 분석
학술지명 한국환경분석학회 저자 윤용배,오은정,김학철,이상태
발표일 2008-05-23

This study included the development of analytical method for determining perchlorate in water sample. The analytical condition was referred in EPA Method 314.0 and 314.1 which use Ion chromatography. Concentrating 10mL raw or treated water sample on to AG16 guard column made it possible to get the LOD(Limit of Detection) of 0.73 ㎍/L. Compared to the concentration method which needed manual operation, the direct Injection method could screen a number of water samples. The LOD of the direct injection method was higher and the sensitivity was lower than those of the concentration method. The RSDs(Relative Standard Deviations) were lower than 2.5 % for peak height and 0.7 % for retention time in preconcentration methods. Matrix elimination could be done using the preconcentration method if perchlorate were in complex matrix of sample. This study also included the development of analytical method for determining perchlorate  using Ion chromatography-Mass Spectrometer(IC/MS/MS) on MRM(Multiful Reaction Monitoring) mode. The LOD value of perchlorate for IC/MS/MS method was 0.005 ㎍/L on MRM mode.

목록