The large size electromagnetic flow-meter was tested to investigate the variation of its error characteristics in the range of low velocity under 0.6m/s using flowmeter calibration system. For the two case of valve opening (100%, 50%), these tests were undertaken three times each for twelve velocity condition from 0.05㎧ to 0.6㎧ with increment of 0.05㎧.
It is shown that the minimum velocity should be higher than 0.25㎧ for φ300㎜ and 0.3㎧ for φ800㎜ to keep manufacturer's(±0.4% of rate). Measurement deviation of flowmeter for φ400㎜ and φ600㎜ was out of expected deviation range suggested by manufacturer, however, its tendency was kept similar to manufacturer's suggestion. In conclusion, error characteristic of electromagnetic flowmeter wasn't changed proportion to its size but depended on the fact that if the flow through the flowmeter is fully developed or not.