Differences between Analytical Results for Perchlorate by Ion Chromatography-Conductivity detection and Ion Chromatography with Tandem Mass Spectrometry Considering EPA Method 314.1 게시글의 제목, 학술지명, 저자, 발행일, 작성내용을 보여줌
Differences between Analytical Results for Perchlorate by Ion Chromatography-Conductivity detection and Ion Chromatography with Tandem Mass Spectrometry Considering EPA Method 314.1
학술지명 Journal of Analytical Chemistry 저자 김학철
발표일 2011-11-30

In this study, an analytical method was developed for determining the perchlorate concentrationsin effluent samples obtained from two refineries. The primary and confirmatory methods usedwere from EPA 314.1. The two methods found similar perchlorate concentrations for the Yrefinery sample; however, there was a significant difference in the concentration values found forthe K refinery sample. The difference was attributed to co-eluting ions with nearly the sameretention time as perchlorate. For more accurate and reliable analysis, the two samples wereanalyzed by Ion Chromatography with Tandem Mass Spectrometry(IC-MS/MS) in MultifulReaction Monitoring(MRM) mode. The results from IC-MS/MS corresponded well to theconfirmatory method. In conclusion, these results show that the primary method of EPA 314.1can return false data due to co-eluting ions with perchlorate.